Operator’s Manual – OPTI CCA-TS2
4-9
4 CALIBRATION AND QUALITY CONTROL
2. If
<Non Secure User ID Entry>
is enabled
in the security settings (see Section 3.2.3), you
will be asked to enter the user ID (Fig. 4-13).
NOTE: Bar-coded user IDs may be entered from
this screen using the bar code scanner.
3. Select the desired level (Fig. 4-14) and press
.
NOTE: If a new lot number of QC material is used,
make sure the ranges have been entered into
the system prior to running a sample. (See
Section 3.2.1.1).
4. Scan the bar-coded strip on the OPTI Cassette
package by holding it 2-3 inches (5-8 cm)
from the bar code scanner located on the
bottom right-hand corner of the analyzer to
automatically record the lot and calibration
information for the specific cassette (Fig. 4-15).
•
The red line from the barcode scanner
should cover the entire bar code.
•
The unit will beep and the status light will
turn green to confirm a valid bar code.
•
In case of an expired cassette, the light will
turn red.
NOTE: Refer to special handling instructions inside
the cassette box for refrigerated cassettes.
NOTE: If the bar code is damaged or unreadable,
press
<Manual Entry>
and enter the bar
code digits printed on the bar code label
using the numeric keypad.
NOTE: A control measurement may be made using
any cassette lot or cassette type.
Fig. 4-14 Select QC Level
Fig. 4-13 Enter User ID
Fig. 4-15 Scan Bar Code