Operating Safety Information
Operating Safety Information
•
Overvoltage Category II when connected to a branch circuit.
•
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules.
Caution:
•
The instrument is designed as a conventional device (closed, not waterproof type).
•
Do not operate the instrument in an explosive environment or in the vicinity of explosive
anesthetic mixtures containing oxygen or nitrous oxide.
•
This instrument is suitable for continuous operation.
•
The power plug is to be plugged into a ground socket only. When using an extension cord,
make sure that it is of the proper size and is properly grounded.
•
Any breakage of the ground lead inside or outside the instrument or a loose ground
connection can cause a hazardous condition when operating the instrument. Intentional
disconnection of the grounding is not permitted.
•
When replacing the fuses, make sure that they are of the same type and rating as the
original fuses. Never use repaired fuses or short-circuit the fuse holders.
This device is a Class 1 Laser product according to the requirements of IEC 60825-1.
The LEDs have been certified as an EXEMPT RISK GROUP in compliance with IEC 62471.
The maximum energy output is as follows:
670 nm (LED): 40 Microwatts max. for 400ms
780 nm (Laser): 40 Microwatts max. for 400ms
850 nm (Laser): 40 Microwatts max. for 400ms
Caution:
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure.
Conditions of Acceptability:
Coin cell battery on main board:
•
Overcharging, short circuiting, reverse charging, mutilation or incineration of the cells must
be avoided to prevent one or more of the following occurrences: release of toxic materials,
release of hydrogen and/or oxygen gas, rise in surface temperature.
•
If a cell has leaked or vented, it should be replaced immediately using protective gloves.
•
If and when necessary, these cells must be replaced with identical new ones from the
same manufacturer. If a cell to be replaced is connected with other cells in series, it is
recommended that the other cells be replaced with new ones at the same time.
•
Reverse polarity installation of the cell in the end product must be avoided.