Operator’s Manual – OPTI CCA-TS2
2-1
2 SETUP
2 SETUP
2.1 Important Safety Instructions
Before you begin installing your OPTI
™
CCA-TS2 Analyzer, carefully read the overview
information in this chapter.
For your own safety and the proper operation of your equipment, always follow these precautions
when working with your OPTI CCA-TS2:
•
Keep the analyzer away from all sources of liquids such as sinks and wash basins.
•
Keep the analyzer away from explosive gases or vapors.
•
Always handle blood samples and collection devices with care.
•
Use approved protective gloves to avoid direct contact with sample.
•
Dispose of OPTI Cassette according to local regulations.
2.2 Choosing a Location
Location is important for trouble-free operation of your analyzer. Before you begin setup, choose
a site that is convenient for your sampling needs and meets the following physical requirements
of the unit:
•
Grounded electrical outlet.
•
Away from direct sunlight.
•
Room temperature within 10 - 30
º
C (50 - 86° F).
•
Maximum relative humidity of 95% (non-condensing).
•
Ample room to allow air to circulate around the unit.
•
Away from strong electromagnetic fields, such as those created by electric motors and
X-ray equipment.
•
Away from explosive gases or vapors.
•
Placed on flat surface with ample room between air vents on bottom of unit and surface to
prevent unit overheating.
NOTE:
Above requirements also apply when the OPTI CCA-TS2 operates on battery
power outside a laboratory setting.