Operator’s Manual – OPTI CCA-TS2
2 SETUP
•
When prompted (Fig. 2-7), insert the gas
bottle in its housing and turn clockwise until
fingertight (Fig. 2-8). Press
.
NOTE: The gas bottle expires 6 months after
installation or after exceeding the
labeled expiration date, whichever
comes first.
NOTE: The bar code contains expiration
information. Four weeks prior to
expiration of the gas bottle, the OPTI
CCA-TS2 will alert the operator once,
as a reminder to order a replacement
gas bottle.
•
When this display appears (Fig. 2-9), press
to install a new gas bottle.
NOTE: If after the initial installation you need
to remove a gas bottle and reinstall
the same bottle, respond
to the
<New Gas Bottle?>
prompt.
The next screen will prompt you to
enter the number of weeks in service
using the numeric keypad (See section
7.5.3). Here you may refer back to the
installation date, which was recorded on
the gas bottle.
The OPTI CCA-TS2 will now begin to warm up
and perform a gas purge, which will be indicated
by a progress bar displayed on the screen.
Once the warm-up is complete, the
<Ready>
display appears (Fig. 2-10).
Fig. 2-8 Insert Gas Bottle
Fig. 2-10 Ready screen
Fig. 2-7 Insert Gas Bottle
Fig. 2-9 New Gas Bottle