9.1 EMC Anomalies
Standard: 61326-1, 300 489-1
Model UWBT-PH:
Conducted RF Immunity over the AC Power adapter Lines
Operating in Bluetooth Mode, AC Power adapter: Tested at 220V/60Hz at 3
Vrms, from 900 KHz to 50 MHz, the pH reading changes from 4.23 to 10.48 pH.
From 50 MHz to 80 MHz, you may experience open pH sensor condition.