Communication Protocol
10
10-10
Field
Values And Description
Byte 23
Sampling interval,
1 - for 10samples/1sec
2 - for 1sample/1sec
3 - for 1sample/10sec
4 - for 1sample/30sec
5 - for 1sample/60sec
Byte 24
For Thermocouple Model, RH/Temp Model and for pH/Temp Model will
have value 0.
For RTD Model has following values possible
1-
American Curve
2-
European Curve
Byte 25
For Thermocouple Model, RTD Model and for RH/Temp Model has
value 0.
And for pH/Temp Model the following value is possible
0
- RTD sensor not present
1
- RTD sensor present
Byte 26 –
Byte 27
For Thermocouple Model, RTD Model and for RH/Temp Model will have
value 0.
And for pH/Temp Model if RTD sensor is not present then these bytes
will have the user set temperature value
Byte 28
Reserved
Byte 29
UWBT internal logging interval,
1
- for 10samples/1sec
2
- for 1sample/1sec
3
- for 1sample/10sec
4
- for 1sample/30sec
5 for 1sample/60sec
Byte 30
Gives the information about the UWBT-L Device internal logging status.
0
– Internal logging OFF
1
- Internal logging ON
For UWBT Device (Basic model), the byte is 0 and this byte will not be
considered in Smartphone.
Byte 31
Log Circular buffer Enable/Disable
0
– Circular buffer Disable
1
– Circular buffer Enable (Default condition
Byte 32 –
Byte 47
Device serial number
These fields contain Alphanumeric values.
Note 1: In pH Model, for example if the user sets the dead band / Low alarm /
High alarm values as 12, the application should send the value as 12.00
Table 10-5. Read Device Settings Frame Field Description Cont.