![background image](http://html1.mh-extra.com/html/omega/rd8800/rd8800_user-manual_4055199055.webp)
Chapter 4 Operation
Page 4-7
The current record status is shown in the Disk Status section in the top left corner of the screen. If the unit
is not recording,
OFF X%
will be displayed where X is the percentage of the disk/card used
.
If the unit is
recording data,
RECORD X%
will be displayed if the unit is set to record manually or
TRIG’D X%
will be
displayed if the unit is triggered to record from an internal alarm or external event. Any disk error will be
shown in a pop-up window.
4.2.3.2 Activate point
This menu item returns temporarily bypassed points to the measurement cycle for measuring inputs and
displaying information.
To Activate a point or points:
From the Function menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
Activate pt
and then press the
ENTER
button. Use the numeric keypad to enter the point number to activate and press the
ENTER
button. The next point is then displayed. Continue to enter the point numbers to activate and then when
point activation is finished, press the
EXIT
button to return to the Function menu.
4.2.3.3 Bypass point
This menu item removes active points from the measurement cycle. A bypassed point will still be in the
database and the message
“
BYPASSED” will be displayed when the bypassed point appears on the
display or is printed.
To Bypass a point or points:
From the Function menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
Bypass pt
and then press the
ENTER
button. Use the numeric keypad to enter the point to bypass and press the
ENTER
button. The
next point is then displayed. Continue to enter the point numbers to bypass and then when finished, press
the
EXIT
button to return to the Function menu.
4.2.3.4 Reset Point
This function allows certain point values to be reset without reprogramming each parameter. The value of
the following types of calculated points may be reset:
Moving average
Resets data to the current value of the Basepoint
High peak
Resets data to the current value of the Basepoint
Low peak
Resets data to the current value of the Basepoint
Totalize
Resets to zero
Only the above listed point types will be prompted in this menu item.
To Reset a point or points:
From the Function menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
Reset pt
and then press the
ENTER
button. Use the numeric keypad to enter the point to reset and press the
ENTER
button. The next
point is then displayed. Continue to enter the point numbers to reset and then when finished, press the
EXIT
button to return to the Function menu.
4.2.3.5 Chart speed
This menu option selects between the Low or High Chart speed. The actual Chart speed is selected in the
PROGram, Chart/Pens, Speed menu option (see section 5.5.1). The default setting for both the Low and
High Chart speed settings is 60 inches per hour.
To select the Chart speed:
From the Function menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
Chart speed
and then press the
ENTER
button. Use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
speed= LO
or
speed= HI
and then press
the
ENTER
button. Press the
EXIT
button to return to the Function menu.