![background image](http://html1.mh-extra.com/html/omega/rd8800/rd8800_user-manual_4055199109.webp)
Page 5-47
Chapter 5 Programming
Refer to the modem manual for other modems. The modem needs to be set up as follows.
Operation
Rockwell etc.
US Robotics, etc
Modem Attention call
AT
AT
Load factory defaults
&F
&F
Ignore DTR line
&D0
&D0
Quiet, no response to DTE
Q1
Q1
Auto answer on the first ring
S0=1
S0=1
Set 9600 baud
S37=9
&U6&N6
No compression
%C0
&K0
No ARQ
n/a
&M0
Use profile 0 on wake up
&Y0
Y0
Store settings into profile 0
&W0
&W0
Set V.34 modulation, no automode, 19200 bps
+MS=11,0,19200
n/a
Set 19200 baud
n/a
&n10
The entire string may not fit in the display window. Use the
LEFT (
←
←
←
←
←
)
or
RIGHT (
→
→
→
→
→
)
button to navigate through
the string. The modem string is then sent once you return to the Program menu or whenever power is applied
to the Recorder. Refer to Chapter 2.4 for modem connection.
5.11 System
The System menu option allows the user control the Beeper, set Passcodes or Alternate Languages, and configure
the Parallel Printer Port.
To modify this System parameters:
From the Program menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
System
and then
press the
ENTER
button. The System menu will be displayed as shown in Figure 5-9.
Use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to select the option you wish to program and then
press the
ENTER
button. The options are described in the following sections.
5.11.1 Beeper
The Beeper menu option allows the user to turn the Beeper On or Off. If On, the
Beeper provides audible feedback to button/key presses, indicates errors and beeps
on alarm condition.
To turn the Beeper On or Off:
From the System menu use the
UP (
↑↑↑↑↑
)
or
DOWN (
↓↓↓↓↓
)
button to highlight
Beeper
and then press the
ENTER
button. The current Beeper status will be displayed as “Beeper=YES*” or “Beeper=NO*”. Press the
YES
button to turn the Beeper On, or the
NO
button
to turn the Beeper Off and the press the
ENTER
to return to the
System menu.
5.11.2 Passcodes
The Passcodes menu option allows the user to program a Passcode to protect the FUNCtion menu and/or the
PROGram and Hidden menus from unauthorized changes. Once a Passcode is set, it must be entered to gain
access to the respective menu. Separate Passcodes can be set for either. Note that if a Passcode is forgotten,
you will not be able to change the configuration of the unit. The only way to change or delete a Passcode is to
know the present Passcode.
Treat Passcodes with respect.
Passcodes
Alt. Language
Printer out
Beeper
Figure 5-4
System Menu
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com