DMTA-10004-01EN, Rev. D, November 2016
Using Special Functions
119
STANDARD
The standard alarm warns you when the actual measured thickness is below a
low reference value or above a high reference value. The reference values are
thickness set points using the current instrument units and resolution.
PREVIOUS THK
To use this function, you must first open a previously stored inspection file. The
alarm uses the previous thickness value as its reference when determining if the
actual thickness is in a high or low alarm condition. The previous thickness alarm
warns you when the actual measured thickness is outside the alarm reference
values when compared to the previous thickness value at each ID location within
the file.
You can choose the
ABSOLUTE DIFF
(differential) mode to set the following
parameters:
— When the thickness differential is smaller than or equal to the
ABSOLUTE
LOSS
value, a low alarm is indicated.
— When the thickness differential is greater than the
ABSOLUTE GROWTH
value, a high alarm is indicated.
Table 8 on page 119 shows the result for an example where both
ABSOLUTE
LOSS
and
ABSOLUTE GROWTH
are set to
1.27 mm
(
0.05 in
.).
You can also choose the
% DIFF
(differential) mode to set the following
parameters:
— When the thickness differential is smaller than or equal to the
% LOSS
value,
a low alarm is indicated.
Table 8 Alarm calculation example with the ABSOLUTE DIFF mode
Previous
values
Actual
thickness
Low alarm High alarm
Calculations
7.62 mm
(0.300 in.)
6.07 mm
(0.239 in.)
Yes
No
7.62–6.07 > 1.27 mm
(0.300–0.239 > 0.050 in.)
7.62 mm
(0.300 in.)
8.03 mm
(0.316 in.)
No
Yes
8.03–7.62 > 0.127 mm
(0.316–0.300 > 0.005 in.)
7.62 mm
(0.300 in.)
7.24 mm
(0.285 in.)
No
No
7.62 – 7.24 < 0.127 mm
(0.300–0.285 < 0.050 in.)
7.62 mm
(0.300 in.)
7.67 mm
(0.302 in.)
No
No
7.67 – 7.62 < 0.127 mm
(0.302–0.300 < 0.005 in.)
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...