DMTA-10004-01EN, Rev. D, November 2016
Chapter 5
74
maximum accuracy is to be achieved, then the coupling technique must be
consistent. In order to accomplish consistent measurements, use a couplant of
reasonably low viscosity; employ only enough couplant to achieve a reasonable
reading; and apply the transducer with uniform pressure. Practice will show the
degree of moderate to firm pressure that produces repeatable readings. In
general, smaller diameter transducers require less coupling force to squeeze out
the excess couplant than larger diameter transducers. In all modes, tilting the
transducer distorts echoes and causes inaccurate readings, as noted below.
Curvature of the test piece
A related issue to this section involves the alignment of the transducer with
respect to the test piece. When measuring on curved surfaces, it is important that
the transducer be placed approximately on the centerline of the part and held as
steadily to the surface as possible. In some cases, a spring-loaded V-block holder
may be helpful for maintaining this alignment. In general, as the radius of
curvature decreases, the size of the transducer should be reduced, and the more
critical transducer alignment will become. For very small radii, an immersion
approach is necessary. In some cases it may be useful to observe the waveform
display as an aid in maintaining optimum alignment. Practice the best way to
hold a transducer with the aid of a waveform display. On curved surfaces, it is
important to use only enough couplant to obtain a reading. Excess couplant forms
a fillet between the transducer and the test surface where sound reverberates and
possibly creates spurious signals that may trigger false readings.
Taper or eccentricity
If the contact surface or back surface of the test piece is tapered or eccentric with
respect to the other, the return echo is distorted due to the variation in sound path
across the width of the beam. The accuracy of the measurement is reduced. In
severe cases, no measurement is possible.
Acoustic properties of the test material
There are several conditions found in certain engineering materials that can
potentially limit the accuracy and range of ultrasonic thickness measurements:
— Sound scattering:
In materials such as cast stainless steel, cast iron, fiberglass, and composites,
sound energy scatters from individual crystallites in the casting or
boundaries of dissimilar materials within the fiberglass or composite.
Porosity in any material can have the same effect. Adjust the instrument
sensitivity to prevent detection of these spurious scatter echoes. This
compensation can in turn limit the ability to discriminate a valid return echo
from the back wall of the material, thereby restricting measurement range.
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...