DMTA-10004-01EN, Rev. D, November 2016
Dual Element Transducer Setups
213
Figure 12
‑
1 The GENERIC SETUP SELECTION screen
3.
Press
[MEAS]
to return to the measurement screen.
4.
Press
[XDCR RECALL]
.
5.
In the menu, select
DEFAULT DUAL ELEMENT
.
6.
In the
DEFAULT DUAL ELEMENT
screen, among the default transducer list,
select the transducer that best matches the frequency and the diameter of the dual
element transducer that you are using (see Table 21 on page 213).
Table 21 Default transducers
Default transducer
Frequency
Tip diameter
D790/791/D799
5 MHz
11.0 mm (0.434 in.)
D792/D793
5 MHz
7.2 mm (0.283 in.)
D794/D795
10 MHz
7.2 mm (0.283 in.)
D797
2 MHz
22.9 mm (0.900 in.)
D798/D7226
7.5 MHz
7.2 mm (0.283 in.)
D7912/D7913
10 MHz
7.5 mm (0.295 in.)
MTD705
5 MHz
5.1 mm (0.200 in.)
D7906 (THRU-COAT)
5 MHz
11.0 mm (0.434 in.)
D7908
7.5 MHz
7.2 mm (0.283 in.)
Generic setups for dual element transducers:
DEF-DE-2MHZ
2 MHz
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...