DMTA-10004-01EN, Rev. D, November 2016
Chapter 7
112
3.
Press
[XDCR RECALL]
.
4.
In the menu, select
DEFAULT HP SINGLE ELEMENT
.
5.
In the
DEFAULT HP SINGLE ELEMENT
screen, highlight the default transducer
setup for the M2008 (
DEFP1
‑
0.5
‑
M2008
) or any custom setup that uses the M2008
transducer.
6.
Press
[MEAS]
to return to the measurement screen with the recalled transducer
setup.
7.
Wipe couplant off the transducer tip.
8.
Press
[2nd F], [CAL ZERO] (Do ZERO)
.
9.
Perform the material sound velocity and zero calibration (see “Calibrating the
Instrument” on page 67).