DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
143
Figure 10
‑
7 B
‑
scan freeze review elements
B
‑
SCAN MAX THK MODE
:
Determines the vertical scale of the B-scan with the following two options:
A
‑
SCAN RANGE
The top and bottom of the vertical scale correspond to the start and end of the
A-scan display range.
SPECIFIED THK
The top of the vertical scale represents the zero value and the bottom of the
scale is the value that you enter in the
B
‑
SCAN MAX THICKNESS
parameter that appears. The allowed range is 0.00 mm to 635.00 mm (0.000 in.
to 25.000 in.).
Olympus does not recommend using the A-scan zoom feature when
B
‑
SCAN MAX
THICKNESS
is set to
A
‑
SCAN RANGE
. The zoom constantly adjusts the start and
the end points of the A-scan range, resulting in a B-scan image where the scale
changes as the thickness changes.
Waveform at minimum
location
Review marker
Minimum location
marker
NOTE
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...