DMTA-10004-01EN, Rev. D, November 2016
Custom Setups for Single Element Transducers
231
The max gain can never be lower then the initial gain and has a maximum value of
99.0 dB.
13.6.2
Initial Gain
The initial gain sets an upper limit on receiver gain in the vicinity of the excitation
pulse (mode 1) or of the interface echo (modes 2 and 3). By effectively making the
excitation pulse or interface echo smaller, the TDG curve allows detection of echoes
occurring close to the pulse. You can adjust the initial gain from 0 dB to the maximum
defined by the
MAX GAIN
setting.
The
INIT GAIN
setting is most critical in applications where minimum thickness
measurements must be optimized. It should always be set with the aid of a reference
standard representing that minimum. In cases where the minimum thickness
capability is less important than penetration and scatter echoes are not a problem, you
can set the initial gain equal to the maximum gain.
The initial gain does the following:
•
Indicates the initial (time-dependent) receiver gain currently selected
•
Amplifies echoes that are close to the Main Bang or interface echoes
•
Starts at time zero and extends to:
— the main bang blank in mode 1
— the end of the interface blank in modes 2 and 3
13.6.3
TDG Slope
The TDG slope controls the rate at which the receiver gain slopes up from the initial
gain level to the maximum gain level. The TDG slope starts at the position of the
MB
BLANK
parameter in mode 1 and at the end of the
IF BLANK
parameter in mode 2
and mode 3. The TDG slope helps suppress reflections from grain structure or fibers.
Generally, you should set the TDG slope as high as possible to reach the maximum
gain as quickly as possible without having the gage hang up on spurious signals. You
can set the slope from 0.0 dB/µs to 39.95 dB/µs.
NOTE
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...