Dual Element Transducer Setups
219
Figure 12
‑
5 Editing point 1 of the V
‑
path calibration
10. Press
[CAL VEL]
.
11. Couple the transducer to the next thickest sample to get a steady thickness
reading, and then press
[ENTER]
.
12. In the
V
‑
PATH CALIBRATION
screen, edit the value for the next point to match
the known thickness.
13. Repeat steps 10 to 12 for the other thicknesses. You need a minimum of three
points and a maximum of ten V-path calibration points.
14. When all known thickness values from all calibration points have been entered,
press
[MEAS]
.
15. At this point you can either press
[CAL VEL]
to view the V-path Calibration table.
OR
Save the V-path calibration as a custom dual element transducer setup:
a)
Press
[SAVE/SEND]
.
b)
In the
SAVE SETUP
screen, enter the desired custom setup file name (see
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...