EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
+7(812) 385-54-28
5
TECHNICAL SPECIFICATIONS
For EMA transducer (EM1401, EM1401 UT)
Measured thickness range for steel
2 to 80 mm
Measured thickness range without gap for steel
2 to 120 mm
Thickness measurement error
0.08 mm
Permissible gap between the sensor and the test object
for ЕМТ14012 transducer
up to 4 mm
Permissible gap between the sensor and the test object
for ЕТМ14013 transducer
up to 6 mm
Permissible skew of the sensor relative to the test object
surface normal
± 25
Minimum permissible radius of curvature of the test
object surface
≥10 mm
Operating frequency
4 MHz
Working ambient temperature range
-20 to +50
C
Operating temperature range of the test object surface
-20 to +80
C
(-20 to +750
C
using the EMT14014Т
transducer*)
Maximum number of measurements per second
16
For PE transducer (EM1401 UT)
Measured thickness range for steel for EDC5P10FS15
transducer
2 to 300 mm
Measured thickness range for steel for EDC5P7.2FS10
transducer
1 to 150 mm
Measured thickness range for steel for EDC10P5F3
transducer
0.5 to 20 mm