EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
+7(812) 385-54-28
21
Thickness measurement algorithm selection
Accumulation Number Selection
By default, the Device uses 128 accumulations of the useful signal and
calculates thickness using the average values. When the Device measures thickness of
a test object with poor surface quality and across a wide gap, it is recommended to
increase the number of accumulations and, thereby, to increase the reliability and
accuracy of the measurements. To enter the menu of accumulation number selection,
press the
button. The accumulation number is selected by the
and
buttons. After selection of the required accumulation number, press the
button.
Accumulation number selection
The accumulation number can be as follows: 8, 16, 32, 64, 128, 256, 512, or
1024;