EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
+7(812) 385-54-28
36
Test object name entering
To enter symbols, letters, and numbers in this window, use the
,
,
and
buttons. Upon completion, press the
button.
For the object of the "Pipe" type, set the necessary number of the measured
points in one ring as well as number of rings along the pipe.
For the object of the "Matrix" type, set the number of the measured points in
horizontal plane and the same in the vertical plane.
Enter these values using the
and
buttons.
The data in the C-scan can be filled both from top to bottom (vertical lines) and
from left to right (horizontal lines). The user can select the order of filling the C-scan
in the "Order" item using the
and
buttons.
The user shall enter the base thickness of the test object and minimum
permissible thickness of the test object. In the course of the C-scan filling, all points
with thicknesses less than the minimum permissible thickness will be displayed in
black. Such points are visible very well on the background. The points with thicknesses
higher that the base thickness will be displayed in green. The points with thicknesses
between the minimum and base values are displayed on the C-scan with gradient.
Upon entering all settings, select item "Start inspection" to start the C-scan
filling and press the
button.