EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
+7(812) 385-54-28
42
EMT14014T
from -20 to
+750
from 2 to
120 mm
from 0 to
3
The transducer has a steel housing and
special heat-resistant protector and is
designed for operation with the objects
heated up to 750
C.
EDC10P5F3
from -5 to
+50
from 0.5
to 20
-
The transducer operates at 10 MHz and
is designed for thickness measurements
of the thin objects and small defects
detection.
EDC5P7.2FS10
from -5 to
+50
from 1 to
150
-
The transducer operates at 5 MHz and is
designed for thickness measurements of
the medium thick objects and suitable
for most of applications.
EDC5P10FS15
from -5 to
+50
from 2 to
300
-
The transducer operates at 5 MHz and is
designed for thickness measurements of
the thick objects and of the objects with
poor surface quality.
Note that, when using the EMT14014T transducer, it cannot be heated to the
temperatures exceeding 300 ºС. Therefore, while operating with the objects heated to
the temperatures exceeding 300 ºС, it is necessary to observe time intervals in
accordance with the table below:
EMT14014T transducer measurement modes
Temperature range, ºС
Measurement duration, seconds
0 – 300
continuously
300 – 700
0 – 5 (time interval between measurements:15
seconds)
Thickness Measurements using the Device
To measure the thickness, do as follows:
1.
Select the transducer to operate with the thickness gauge as described in
the previous item.
2.
If necessary, clean the surface and apply contact liquid (when using the
piezoelectric transducers).