EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
+7(812) 385-54-28
41
Brightness
This item enables to set the percentage screen brightness. To change the value,
press the
and
buttons.
Restore default
This item enables to restore all defaults of the Device. To select this menu item,
press on the
button.
About
This item enables to know the software version. To select this menu item, press
on the
button.
Transducer Selection
The transducer is selected based on the test object temperature, base thickness,
surface quality, etc. The characteristics of all applicable sensors are given in the
Comparative table provide below:
Comparative table for transducer selection
Transducer
Test object
temperature
range,
C
Thickness
range,
mm
Operating
gap, mm
Description
EMT14012
from -20 to
+80
from 2 to
120 mm
from 0 to
4
Compact transducer, suitable for most of
applications. It is supplied together with
the Device as standard scope of supply.
EMT14013
from -20 to
+80
from 2 to
120 mm
from 0 to
6
The transducer has the enhanced
magnetic system and is used in such
cases when operation with the maximum
operating gap or on the objects with poor
surface quality due to heavy corrosion is
required.