Chapter 2
Microscopy Operations
48
Ep
i-fluo
resce
n
ce
Micro
sco
p
y
Preparation Focusing
Microscopy
Epi-fluorescence
Microscopy Procedure
1
2
3
4
5
6
7
8
9 10 11 12 13 14
■
■
■
■
■
Concurrent Use of Epi-Fluorescence/Phase Contrast Microscopy
By bringing the following optical elements into the optical
path, epi-fluorescence and phase contrast can be
simultaneously observed by microscope:
Filter cube for the epi-fluorescence cube turret
Ph objective
PH module for the condenser
•
If the diascopic image is not shown under
epi-fluorescence microscopy, press the
dia-illumination ON/OFF switch to turn on the lamp.
•
Remove the GIF filter if it is installed on the field
lens.
IN
OUT
1
2
A
λ
F. STO
P
A
EX.
2
3
4
5
6
Concurrent use of epi-fluorescence/phase contrast
microscopy
•
Adjust the fluorescent image brightness with ND
filters in the epi-fluorescence attachment, and
phase contrast image brightness using ND filters in
the main body. Dia-illumination should be
sufficiently dimmed using the ND filter.
ND4
ND8
ND16
Brightness adjustment of fluorescent image
ND
8
ND
32
OUT
IN
NCB
11
Adjusting brightness of phase contrast image
Ph objective
Filter cube
PH module
Summary of Contents for ECLIPSE Ni-E
Page 2: ......
Page 24: ...Chapter 1 Components 10 Components ...
Page 29: ...Chapter 2 Microscopy Operations 15 Before Microscopy Operation Flowchart ...
Page 30: ...Chapter 2 Microscopy Operations 16 Before Microscopy Operation Flowchart ...
Page 31: ...Chapter 2 Microscopy Operations 17 Before Microscopy Operation Flowchart ...