4.10 Response Calibration
BP4620
4-79
<9> Press the
ENTER
key to switch on the response calibrator.
The [Adjusted characteristics] can be selected through the above operation.
<10> Select SQU (square wave) as the waveform to be output.
For the waveform setting procedure,
refer to “4.3.6 Setting of waveform”
.
<11> Switch on the output.
For the output on switching procedure,
refer to “4.3.2 Output on/off”
.
<12> Adjust the three adjustment knobs towards the rear of the operation panel while observing
the waveform with an external oscilloscope.
<13> Once the adjustments have been completed, set the operation panel back to its original
position.
(
Refer to “3.2 Tilting LCD Screen”.)
The response characteristics can be improved using the adjustment knobs, but an unstable status
(oscillation status) may also result depending on how the adjustment is performed according to the
load conditions (including no-load). Therefore, when using adjustment characteristics, make sure that
the adjustment knobs are set to their slowest and most stable settings on the left (counterclockwise)
before using an external oscilloscope for waveform monitoring and before adjustment.
!
CAUTION
Summary of Contents for BP4620
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual nbn Austria GmbH...
Page 2: ...DA00016125 006 BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual...
Page 3: ......
Page 15: ......
Page 16: ...BP4620 1 1 1 OVERVIEW 1 1 General 1 2 1 2 Features 1 2 1 3 Principle of Operation 1 4...
Page 21: ......
Page 39: ......
Page 165: ...6 1 Error Messages BP4620 6 4 Figure 6 1 Protective Function Related Error Display Examples...
Page 171: ......
Page 187: ......
Page 206: ...BP4620 9 1 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 2...
Page 211: ......
Page 213: ......