7.8 Performance Test
BP4620
7-14
Amplitude accuracy test of CC mode internal DC signal source
Load: Short via a current shunt resistor
Measurement connection: Measure the detection voltage of the current shunt resistor with a DC
voltmeter and convert the measured value to a current value.
Settings: After setting initialization has been performed, set the operation mode and DC output
current according to the following table.
Judgment: If the output current of the BP4620 falls within the rated ranges in the following table,
the operation is judged to be normal.
Operation Mode Setting
DC Output Current Setting
Rated Range
CC-INT
+20 A
+19.84 to +20.16 A
20 A
20.16 to
19.84 A
Frequency accuracy test of internal AC signal source
Load: 3
resistance load
Measurement connection: Measure the signal (1 Vp-p) of the voltage monitor output terminal of
the BP4620 with a frequency counter.
Settings: After setting initialization has been performed, set the frequency and AC output
voltage according to the following table.
Judgment: If the output frequency of the voltage monitor of the BP4620 falls within the rated
range in the following table, the operation is judged to be normal.
Frequency Setting
AC Output Voltage Setting
Rated Range
10000 Hz
100 Vp-p
9999 to 10001 Hz
Amplitude accuracy test of CV mode internal AC signal source
Load: 3
resistance load
Measurement connection: Measure the voltage of the output terminal of the BP4620 with an AC
voltmeter.
Settings: After setting initialization has been performed, set the frequency and AC output
voltage according to the following table.
Judgment: If the output voltage of the BP4620 falls within the rated ranges in the following
table, the operation is judged to be normal.
Frequency Setting
AC Output Voltage Setting
Rated Range
500 Hz
120 Vp-p
119.2 to 120.8 Vp-p
(42.15 to 42.70 Vrms)
10 Vp-p
9.75 to 10.25 Vp-p
(3.446 to 3.623 Vrms)
Amplitude accuracy test of CC mode internal AC signal source
Load: Short via a current shunt resistor
Measurement connection: Measure the detection voltage of the current shunt resistor with an AC
voltmeter, and convert the measured value to a current value.
Settings: After setting initialization has been performed, set the operation mode, frequency, and
AC output current according to the following table.
Judgment: If the output current of the BP4620 falls within the rated ranges in the following table,
the operation is judged to be normal.
Operation Mode Setting
Frequency Setting
AC Output Current Setting
Rated Range
CC-INT
500 Hz
60 Ap-p
59.64 to 60.36 Ap-p
(21.1 to 21.34 Arms)
6 Ap-p
5.91 to 6.09 A
(2.09 to 2.152 Arms
)
Summary of Contents for BP4620
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual nbn Austria GmbH...
Page 2: ...DA00016125 006 BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual...
Page 3: ......
Page 15: ......
Page 16: ...BP4620 1 1 1 OVERVIEW 1 1 General 1 2 1 2 Features 1 2 1 3 Principle of Operation 1 4...
Page 21: ......
Page 39: ......
Page 165: ...6 1 Error Messages BP4620 6 4 Figure 6 1 Protective Function Related Error Display Examples...
Page 171: ......
Page 187: ......
Page 206: ...BP4620 9 1 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 2...
Page 211: ......
Page 213: ......