4.10 Response Calibration
BP4620
4-82
(2) CC mode/inductive load
Basically, adjustments can be performed with the T and I adjustment knobs.
Turning the V adjustment knob to the right lowers the ringing frequency and makes it easier to
suppress ringing with the I adjustment knob, but normally the BP4620 is used with the V
adjustment knob turned all the way to the left.
The operations when the adjustment knobs are turned are described below.
If T adjustment knob is turned to the right
The rise/fall become faster, but the ringing frequency also rises, and its amplitude increases.
If I adjustment knob is turned to the right
The ringing amplitude is decreased but the rise/fall become slower.
To make the rise/fall as fast as possible and suppress ringing, perform adjustments by repeating
the following operations.
<1> Speed up the rise/fall with the T adjustment knob.
<2> Suppress ringing when it occurs with the I adjustment knob.
Table 4-25 shows adjustment examples for the three adjustment knobs, and Figures 4-26 to 4-28 show
output current waveform examples for each case.
Table 4-25. Adjustment Example for CC Mode/Inductive Load
L
Calibration Position
T V I
150
H 5
1
7
390
H 3
1
7
1.6 mH
1
1
7
Summary of Contents for BP4620
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual nbn Austria GmbH...
Page 2: ...DA00016125 006 BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual...
Page 3: ......
Page 15: ......
Page 16: ...BP4620 1 1 1 OVERVIEW 1 1 General 1 2 1 2 Features 1 2 1 3 Principle of Operation 1 4...
Page 21: ......
Page 39: ......
Page 165: ...6 1 Error Messages BP4620 6 4 Figure 6 1 Protective Function Related Error Display Examples...
Page 171: ......
Page 187: ......
Page 206: ...BP4620 9 1 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 2...
Page 211: ......
Page 213: ......