Chapter 5 Ion Measurement (F-53,55)
5.4 Changing Measurement SETUP Conditions
86
HORIBA
●
Setting alarms
The alarm function detects the measured value out of the upper or lower limit value, and
either displays them on the screen or outputs them from the external output terminals.
The values out of the limit value are displayed in the different color on the screen.
1.
In the ION SETUP screen, press
↑
or
↓
key to select the LIMIT
option. Press
←
or
→
key to select
OFF or ON.
Selecting OFF:
1.
Press F4 (RETURN).
Selecting ON:
1.
Press the ENTER key.
The cursor moves to the upper limit value setting part.
2.
Press the ENTER key again.
The display color of the setting part changes, enabling the input by pressing
↑
or
↓
key.
Setting range: 0.0
μ
g/L to 999 g/L(0.00
μ
mol/L to 999 mol/L)
Press
←
or
→
key to change the auxiliary unit.
3.
When the setting is completed, press the ENTER key.
The setting is determined, and the display color of the setting part changes.
4.
Press the lower limit item by
↓
key.
5.
Press the ENTER key.
The display color of the setting part changes, enabling the input by pressing
↑
or
↓
key.
Setting range: 0.0
μ
g/L to 999 g/L(0.00
μ
mol/L to 999 mol/L)
Press
←
or
→
key to change the auxiliary unit.
6.
Press the ENTER key to determine the setting.
The display color of the setting part
changes.
7.
Press F4 (RETURN).
The setting is reflected, and the display returns to the Measurement screen.
The set value automatically selects the appropriate auxiliary unit.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...