Chapter 5 Ion Measurement (F-53,55)
5.2 Measurement Using a Calibration Curve
64
HORIBA
Repeatability check
To confirm the measurement accuracy in the Ion measurement, the repeatability check is
available.
1.
After the calibration, press F4 (REPEATABILITY) in CAL screen.
2.
Immerse the electrode in the standard solution calibrated by the first point, and
press the CAL key.
The difference between the calibrated value and measured value is displayed.
When the repeatability check is successfully completed, the electrode status “OK”
appears on the screen.
●
Function key at Ion calibration
●
CHECK:
To perform periodical check, refer to "
●
CAL DATA:
To review the former calibration status, refer to "5.5 Check Display" page 89
●
CAL CLEAR:
To clear all the calibrated values
For the repeatability setup, refer to "
Potential, Concentration and Temperature
at calibration
Standard solution number for the check
result calculation
Potential, Concentration and Temperature
at repeatability check
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...