Chapter 3 pH MEASUREMENT
3.4 Calibration Setting
54
HORIBA
Sample printout of the result data
For accurate measurement" page 52
When electrodes are defective" page 52
●
SIMULATOR mode
Input the results given by the optional pH checker, X-51 to the meter, and perform functional
check.
For detailed operation, refer to the instruction manual for the pH checker X-51.
1.
Set the resolution to 0.001 pH in the pH SETUP screen.
2.
Set the temperature compensation to ATC.
3.
According to the operation guide, perform functional check as calibration by the
standard solutions of pH7, pH4 in the following order.
①
Calibration by standard solutions:
pH6.865
pH4.008
②
Linearity check
pH0.000
pH4.000
pH7.000
pH10.000
pH14.000
③
Indication check by inputting high impedance
Input pH0.000
Input pH14.000
④
Temperature indication check
The displayed pH values at this time have nothing to do with the check result.
0.0
℃
30.0
℃
60.0
℃
100.0
℃
When the calibration range error occurs at step
①
above, the screen displays the error and
the check stops.
OK/CHECK of judgment
Date of check
Asymmetry potential
Sensitivity
Calibrated value for each standard solution
Repeatability
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...