Chapter 6 CONDUCTIVITY MEASUREMENT(F-54,55)
6.2 Cell Constant Setting
92
HORIBA
6.2
Cell Constant Setting
The cell constant can be input.
1.
Press the CAL key in the Measure-
ment screen.
A window appears, and the COND
cell constant can be input.
2.
Press
↑
or
↓
key and set the cell
constant displayed on the elec-
trode.
The unit indication of the cell constant
depends on the electrode. Convert the unit to
the one for the meter before input.
100 m
-1
= 1 cm
-1
1000 m
-1
= 10 cm
-1
10 m
-1
= 0.1 m
-1
3.
Press
←
or
→
key and set the
digit.
4.
Press the MEAS key.
The Measurement screen appears. The
display shows the instantaneous mea-
surement value by the set cell con-
stant.
To cancel, press F1 (CANCEL).