16.2 Measuring mV (oxidation-reduction potential
[ORP])
●
ORP principles
ORP (or Oxidation-Reduction Potentialî) is an abbreviation for oxidation-reduction potential.
ORP is the energy level (potential) determined according to the state of equilibrium between
the oxidants and reductants that coexist within a solution.
One type of equilibrium in a solution
••••••••••
①
If only
①
exists within a solution, a metal electrode (platinum, gold, etc.) and a reference
electrode are inserted into the solution, forming the ORP measuring system shown in Fig.1.
Measuring the potential (ORP) that exists between the two electrodes enables the potential to
generally be expressed by the following equation.
••••••••••
②
E: Electric potential E: Coefficient R: Gas coefficient
T: Absolute temperature n: Electron count
F: Faraday constant a: Activity
Fig.1 ORP measuring system
For example, for a solution in which trivalent iron ions coexist with bivalent iron ions, equa-
tions
①
and
②
would be as follows.
••••••••••••
①
'
••••••••••••
②
'
When only one type of state of equilibrium
①
exists in the solution, the ORP of the solution
can only be determined by equation
②
. What is important here is that ORP is determined by
the ratio of activity between the oxidant and the reductant ) (using the equation ). In actuality,
however many kinds of states of equilibrium exist simultaneously between various kinds of
ions, in most solutions. This means that under actual circumstances, ORP cannot be expressed
using the simple equation shown above and that the physical and chemical significance with
respect to the solution is not very clear.
E
E
0
RT
nF
------
aM
Z n
–
(
)
+
aM
Z
+
--------------------------
ln
–
=
Potentiometer
E
E
0
RT
F
------
aFe
2
+
aFe
3
+
-------------------
ln
–
=
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...