Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
F-52, 53, 54, 55
207
●
ERR No.15 Ethernet error
Definition
Connection of the Ethernet is abnormal.
●
ERR No.16 Ethernet timeout
Definition
Abnormality occurs in Ethernet during communication.
Cause
How to solve problem
Faulty Ethernet con-
nection or setting
Check the connection and setting of the net-
work. Check the PC software.
Cause
How to solve problem
Faulty Ethernet con-
nection or setting dur-
ing communication
Check the connection and setting of the net-
work. Check the PC software.