Chapter 16 Reference
16.6 Default settings
F-52, 53, 54, 55
249
ION SETUP
ELECTRODE
MODEL
○
Selection formula from
attached material or freed
input (9 characters)
ION SETUP
ELECTRODE LOT
○
7 digit integer
ION SETUP
LIMIT
○
OFF
OFF, ON(0.00
μ
g/L to 999g/L)
COND SETUP
UNIT
○
S/m
S/m, S/cm
COND SETUP
TEMP
COMPENSATION
○
AUTO
AUTO, MANUAL
(0.0
℃
to 100.0
℃
)
COND SETUP
TEMP
COEFFICEINT
○
ON(2.00%/
℃)
OFF, ON
(0.00 to 100.00%/
℃
),
PURE WATER
COND SETUP
ELECTRODE
MODEL
○
Selection formula from
attached material or freed
input (9 characters)
COND SETUP
ELECTRODE LOT
○
7 digit integer
COND SETUP
LIMIT
○
OFF
OFF, ON
(0.0
μ
S/m to 199.9S/m)
SALT RESIS-
TANCE SETUP
SALT LIMIT
○
OFF
OFF, ON(0.00% to 4.00%)
SALT RESIS-
TANCE SETUP
RESISTANCE
LIMIT
○
OFF
OFF, ON
(0.0
Ω
•m to 199.9M
Ω
•m)
SYSTEM SETUP DISPLAY
○
DIGITAL
DIGITAL, GRAPH, ANALOG
SYSTEM SETUP 2CH DISPLAY
○
ON
ON,OFF
SYSTEM SETUP AUTO HOLD
○
NORMAL
EXACT, NORMAL, BRIEF,
TIME(2 to 999 seconds), OFF
SYSTEM SETUP INTERVAL MEM-
ORY
○
OFF
OFF, ON(1 to 999 seconds)
SYSTEM SETUP PRINTER OUT-
PUT
○
BRIEF
BRIEF, EXACT
SYSTEM SETUP LANGUAGE
○
ENGLISH
JAPANESE, ENGLISH
SYSTEM SETUP SOUND
○
ON
ON, OFF
DATE AND TIME
SETUP
DATE AND TIME
SETUP
Jun 01 2003
Jun 01 2000 to 2099
MAINTENANCE
BRIGHTNESS OF
DISPLAY
Middle
26 step gradation
MAINTENANCE
DISPLAY COLOR
0
8 steps
MAINTENANCE
CHARACTER
OFF
OFF, ON
MAINTENANCE
BACK LIGHT OFF
TIME
30 minutes
0 to 999 minutes
*Screen does not shift to
power saving mode if *0 min.
is set here.
Category
Item
○
: shows items to
store setting on opera-
tor basis.
Default values
(Set value on
shipment from
factory)
Setting range (Setting item)
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...