Chapter 12 RS-232C Communication
12.4 List of Serial Command
F-52, 53, 54, 55
157
●
On-Line Operation Commands(2)
Format
Header, Item code, Parameter <CR><LF>
Command
for...
Command
Parameter
Description of
parameter
Note
Header
Comma
nd code
Start mea-
surement
C
MS
x
x: channel number
Measurement starts on the channel
which is in the current mode.
Start pH
standard
solution
calibration
C
CP
xxxxxx
xxxxxx: Value to
specify for other
buffers: 6 digits
including decimal
point (up to 3 deci-
mal places)
(
When the stan-
dard solution is
customized)
Effective in the pH measurement ready
condition or in the HOLD condition.
Calibration is executed by other buffer
specify value when “Custom” is selected
for the standard solution and by the
specified standard solution value in
other cases.
In the modes other than “Custom”, cali-
bration starts regardless of the value in
other buffer.
When “
Check before use
” is selected,
the check before use is executed if this
command is issued after calibrating
specified points.
In other cases, calibration can be per-
formed exceeding the calibration points.
However, after calibration of five points
is complete, the error message (Incor-
rect operation) is responded for the suc-
ceeding command.
Start ion
calibration
C
CI
xxxx,y
xxxx: standard
solution concentra-
tion
3-digit effective
number including
decimal point
The available
range for input:
0.00 to 999
y: Auxiliary unit of
the standard solu-
tion
Effective in the ion measurement ready
condition or in the HOLD condition.
Calibration can be performed exceeding
the calibration points. However, after
calibration of 5 points is complete, the
error message (Incorrect operation) is
responded for the succeeding com-
mand.
Start
conductivity
cell constant
calibration
C
CD
xxxxx,y
xxxxx: Conductivity
value
y: Auxiliary value
for the conductiv-
ity value
Start salinity
calibration
C
CS
xxxxx
xxxxx:
Concentration of
the standard solu-
tion
Clear
calibration
C
CC
x
x: channel number
The calibration value of the channel
being displayed currently will be
cleared.
Data clear
C
DC
This command clears the data retained
in the memory.
Data IN
specification
C
IN
Start the interval memory when the
interval memory is set to ON.
When set to OFF, manual data memory
is applied.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...