Chapter 7 SALINITY MEASUREMENT (F-54,55)
7.1 Operation of Salinity Measurement
104
HORIBA
Chapter 7
SALINITY MEASUREMENT
(F-54,55)
The salinity concentration is calculated from the measured value of conductivity. This is why
when the cell constant is set by conductivity measurement, there is no need to input the cell
constant. Cell constant needs input only when it is not set. In this case, set the cell constant by
referring to "6.2 Cell Constant Setting" page 92.
7.1
Operation of Salinity Measurement
In any Measurement screen for
CH2, press the MODE key to pro-
ceed to the SALT Measurement
screen.
The displayed screen shows
the display of salinity instantaneous
measurement.
For F-54 and F-55, press F4 (SELECT CH)
to select CH2.
1.
Press the MEAS key.
The HOLD measurement starts.
When the indication stabilizes, the
HOLD indication lights up and the
measured value is displayed.
Blinking
Lit