Chapter 7 SALINITY MEASUREMENT (F-54,55)
7.2 Salinity Calibration
F-52, 53, 54, 55
105
7.2
Salinity Calibration
The salinity conversion is available by con-
ductivity, yet the calibration using the stan-
dard solution of sea water can obtain an
accurate salinity concentration.
Perform the calibration in the temperature
mentioned in the standard sea water as in the
following step:
1.
Immerse the electrode in the stan-
dard sea water.
2.
In the SALT Measurement screen,
press the CAL key.
The SALT CALIBRATION screen
appears.
3.
Press
↑
or
↓
key to set the con-
centration of standard solution
used for the calibration.
Setting range: 0.00 to 5.00%
4.
Press the CAL key.
The calibration starts.
When the calibration is completed, the
selected Salt Calibration Coefficient is
set.
5.
Press the MEAS key to return to
the Measurement screen.
●
Calibration data display
1.
In the SALT CALIBRATION
screen, press F3 (CAL DATA).
The calibration data is displayed.
Lit
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...