Chapter 6 CONDUCTIVITY MEASUREMENT(F-54,55)
6.4 Measurement Setting Change
96
HORIBA
6.4.3 Temperature conversion coefficient
The conductivity of the solution changes according to the solution temperature. To compare
the solution conductivity, it is necessary either to perform the measurement in a certain tem-
perature or to measure the solution by converting the temperature to the standard temperature.
In the measurement of ACCURATE, the measurement solution should be kept in the standard
temperature typically by a thermostatic chamber.
1.
In the COND SETUP screen, press
↑
or
↓
key to select the TEMP
COEFICIENT option, and press
←
or
→
key to select OFF, ON or
PURE WATER.
In setting to PURE WATER, refer to “Pure
water temperature conversion (double tem-
perature compensation)” described in JIS
K0130.
2.
In selecting ON, pressing the
ENTER key moves the cursor to
the coefficient setting part. Press-
ing the ENTER key again changes
the display color, enabling the
value input by
↑
or
↓
key.
Setting range: 0.00 to 10.00 %/
℃
3.
Press the ENTER key.
The setting is determined, and the dis-
play color of the setting part changes.
4.
Press F4 (RETURN).
The setting is reflected, and the dis-
play returns to the Measurement
screen.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...