Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
204
HORIBA
●
ERR No.06 Abnormal calibration point
Definition
6th point calibration was tried.
There is no influence to the calibration data up to the 5th point.
Refer to "3.4.3 Calibration point setting" page 46.
●
ERR No.07 The pH meter cannot identify the standard buffer
Definition
If the automatic standard-buffer identifica-
tion function of the meter cannot identify
kinds of standard buffer, re-calibrate the
meter after performing the appropriate mea-
sures, below.
Cause
How to solve problem
6th point calibration
was executed.
Regulate the calibration up to 5 points.
This error message is released by setting the
pH meter to measurement mode.
Cause
How to solve problem
There is a problem
with the standard solu-
tion.
Prepare the standard solution anew.
There is a problem
with the standard solu-
tion setting.
Check the kind of the calibration solution of
the NIST, USA standard setting and the one
used for calibration and match them.
The responsive mem-
brane is dry or dirty.
Measure after cleaning the responsive mem-
brane and soaking it in pure water for 24
hours.
Reference internal
solution is poor or
dirty.
Replace the standard solution with new solu-
tion.
The responsive mem-
brane is damaged or
worn out.
Replace the electrode.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...