Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
206
HORIBA
●
ERR No.11 Cell constant out of range
Definition
The cell constant is out of the range of 0.7 to
1.3. Delete data stored in the memory con-
firming their contents.
●
RR No.12 CF Write over defective
Definition
The compact flash card or its contents are abnormal.
●
ERR No.13 CF over capacity
Definition
The capacity of the Compact flash memory is exceeded.
●
ERR No.14 CF not inserted
Definition
The compact flash card is not correctly inserted.
Cause
How to solve problem
COND electrode is at
the end of the useful
life.
Replace the electrode.
Improper standard
solution
Prepare standard solution.
Cause
How to solve problem
Compact flash is not
formatted.
Format Compact flash.
I/F of Compact flash
card cannot be read.
File system is abnor-
mal.
Prepare a new Compact flash.
Cause
How to solve problem
Capacity of Compact
flash memory is
exceeded.
Delete data.
Prepare a new Compact flash.
Cause
How to solve problem
Compact flash is not
inserted.
Confirm that the Compact flash is correctly
inserted.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...