Chapter 9 DATA MEMORY
9.1 Setting Memory Type
F-52, 53, 54, 55
113
Chapter 9
DATA MEMORY
The measurement data can be stored and at the same time retrieved.
There are two ways to store data into memory: pressing “Data Memory” of the function key at
any time in the Measurement mode (Manual Memory) and automatically memorizing by a
certain interval (Interval Memory).
9.1
Setting Memory Type
Select either Interval memory or Manual
memory.
1.
Turn the power ON.
After the startup screen, the SETUP
screen appears.
Pressing ESC key in the Measurement
screen also shows the SETUP screen.
2.
Pressing
↑
or
↓
key, select SYS-
TEM SETUP option, and press
ENTER key.
The SYSTEM SETUP screen appears.
3.
Pressing
↑
or
↓
key, and select
INTERVAL MEMORY.
4.
Press
←
or
→
key and select ON
or OFF.
When set to ON: go to "
memory" page 114.
When set to OFF: go to "
●
Manual memory
1.
To use Manual memory function, press
←
or
→
key and select OFF.
2.
Press F4 (RETURN).
The setting is reflected, and the display returns to the SETUP screen.
3.
Press MEAS key.
The display returns to the pH Measurement screen.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...