NI 5402/5406 Calibration Procedure
42
ni.com
4.
Set the analog path by calling the niFgen Property Node and selecting
Output Attributes»Analog Path
.
5.
Set the gain DAC value by calling the niFgen Property Node and
selecting
Calibration»Gain DAC Value
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_ATTR_
ANALOG_PATH
value
:
NIFGEN_VAL_
FIXED_LOW_GAIN_
ANALOG_PATH
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_GAIN_DAC_VALUE
value
:
1,700