NI 5402/5406 Calibration Procedure
70
ni.com
b.
Set the main DAC value by calling the niFgen Write Binary 16
Analog Static Value VI.
c.
Commit the attribute values to the device by calling the niFgen
Commit VI.
d.
Wait 500 ms for the output to settle.
e.
Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
measurement 0
, which is used in step 6.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
value
:
25,233
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal