National Instruments Corporation
73
NI 5402/5406 Calibration Procedure
Adjusting the Main Path Post-Amplifier Gain and Offset
1.
Set the main DAC value by calling niFgen Write Binary 16 Analog
Static Value VI.
2.
Set the waveform to user-defined by calling the niFgen Property Node
and selecting
Standard Function Output»Waveform
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
value
:
0
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_FUNC_WAVEFORM
value
:
NIFGEN_VAL_
WVFM_USER