National Instruments Corporation
79
NI 5402/5406 Calibration Procedure
9.
Adjust the post-amplifier main path gain and offset by calling the
niFgen Cal Adjust Main Path Post Amp Gain And Offset VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Cal
AdjustMainPath
PostAmpGainAnd
Offset
using the
following parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
configuration
: The
Current Configuration
value for the current
iteration from Table 9
mainDACValues
:
{
0
,
0
}
gainDACValues
:
{
2000
}
offsetDACValues
:
{
38000
,
27000
}
measuredOutputs
:
{
measurement 0
,
measurement 1
}