NI 5402/5406 Calibration Procedure
14. Close the instrument driver session and save the calibration date and
temperature by calling the niFgen Close Ext Cal VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_CloseExtCal
using the following
parameters:
vi
: The session handle returned
from
niFgen_InitExtCal
.
action
: If the external
adjustment procedure
completed without any errors,
use
NIFGEN_VAL_EXT_CAL_
COMMIT
. This function then
stores the new calibration
constants, updated calibration
dates, updated calibration
temperatures in the onboard
EEPROM.
If any errors occurred during
the external adjustment
procedure, or if you want to
abort the operation, use
NIFGEN_VAL_EXT_CAL_
ABORT
. This function then
discards the new calibration
constants and does not change
any of the calibration data
stored in the onboard
EEPROM.