National Instruments Corporation
93
NI 5402/5406 Calibration Procedure
c.
Initiate waveform generation by calling the niFgen Initiate
Generation VI.
d.
Use the frequency counter to measure the frequency of the
generated waveform. This value is the new measured frequency.
e.
Repeat steps 7a through 7d until the difference between the
measured frequency and the desired frequency (10 MHz) is less
than or equal to the tolerance (4.5 ppm).
8.
Abort waveform generation by calling the niFgen Abort
Generation VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal
.