NI 5402/5406 Calibration Procedure
48
ni.com
17. Enable the analog output by calling the niFgen Property Node and
selecting
Basic Operation»Output Enabled
.
18. Commit the attribute values to the device by calling the niFgen
Commit VI.
19. Wait 500 ms for the output to settle.
20. Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
external measurement 0
, which is used in
step 30.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_OUTPUT_ENABLED
value
:
VI_TRUE
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal
.