NI 5402/5406 Calibration Procedure
40
ni.com
The following figure shows the programming flow for an external
calibration procedure.
Figure 3.
NI 5402/5406 External Calibration Procedure
Initialize Analog Output Calibration
Adjust Main Path Pre-Amp Offset
Adjust Main Path Pre-Amp Gain
Adjust Main Path Post-Amp
Gain and Offset
Adjust Oscillator Frequency
Adjust Flatness Correction
Adjust Calibration ADC