NI 5402/5406 Calibration Procedure
64
ni.com
c.
Commit the attribute values to the device by calling the niFgen
Commit VI.
d.
Wait 500 ms for the output to settle.
e.
Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
measurement
0
, which is used in step 5.
f.
Set the gain DAC value by calling the niFgen Property Node and
selecting
Calibration»Gain DAC Value
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_GAIN_DAC_VALUE
value
:
1,000