NI 5402/5406 Calibration Procedure
76
ni.com
6.
Set the analog path by calling the niFgen Property Node and selecting
Output Attributes»Analog Path
.
7.
Set the post-amplifier attenuation by calling the niFgen Property Node
and selecting
Calibration»Post-Amplifier Attenuation
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_ATTR_
ANALOG_PATH
value
: The
Analog Path
value for the current
iteration from Table 9
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_POST_AMPLIFIER
_ATTENUATION
value
: The
Post-Amplifier
Attenuation
value for the
current iteration from
Table 9