NI 5402/5406 Calibration Procedure
24
ni.com
7.
Configure the waveform by calling the niFgen Configure Standard
Waveform VI.
8.
Create an array of waveform samples for the positive full-scale DC
waveform. This array should contain 16,384 samples with each sample
having the value
1.0
(representation: double).
9.
Define the waveform by calling the niFgen Define User Standard
Waveform VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_
ConfigureStandard
Waveform
using the
following parameters:
vi
: The session handle
returned from
niFgen_init
.
channelName
:
"0"
waveform
:
NIFGEN_VAL_WFM_USER
frequency
:
1,000,000
amplitude
:
1
startPhase
:
0
dcOffset
:
0
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_
DefineUserStandard
Waveform
using the
following parameters:
vi
: The session handle
returned from
niFgen_init
.
channelName
:
"0"
waveformSize
:
16,384
waveformDataArray
:
The array of waveform
samples created in step 8