NI 5402/5406 Calibration Procedure
82
ni.com
4.
Set the flatness correction state by calling the niFgen Property Node
and selecting
Calibration»Flatness Correction Enabled
.
5.
Configure the waveform by calling the niFgen Configure Standard
Waveform VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi
: The session handle
returned from
niFgen_
InitExtCal
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_FLATNESS_
CORRECTION_ENABLED
v
alue
:
VI_FALSE
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_
ConfigureStandard
Waveform
using the
following parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
waveform
:
NIFGEN_VAL_WFM_SINE
frequency
:
50,000
amplitude
:
2.00
startPhase
:
0
dcOffset
:
0